Open original FDA record ↗Recall record
- FDA identifier
- Z-1740-2014
- Company in FDA record
- Ivoclar Vivadent, Inc.
- FDA classification date
- Jun 9, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A stock check revealed broken glass fibers between the connector and the fiber bundle in a large number of these light probes.
Source facts
- Action
- lvoclar Vivadent, Inc. sells the affected medical device through Dental Dealer partners (initial consignees) who in turn sell directly to the dentist's office. In this case, lvoclar Vivadent, Inc. handled direct drop shipment of the device to end user dental offices in 80% of the cases. Where lvoclar Vivadent has direct shipment information available, dentist offices will be notified directly. The Dental Dealer partners will be notified of this recall and asked to cooperate and provide name and address listings for all dental office customers where they have shipped the Pin Point probe. The dealer will also be given a customer communication to send directly to the dentists purchasing the device.
- Initiated
- 2014-04-14
- Root cause
- Other
- Terminated
- 2015-11-05
- Distribution
- Worldwide Distribution: US Nationwide and the country of Canada.
- Recall number
- Z-1740-2014
- Recall status
- Terminated
- CDRH record ID
- 126760
- FDA classified
- 2014-06-09
- Recall event number
- 67998