Open original FDA record ↗Recall record
- FDA identifier
- Z-1740-2024
- Company in FDA record
- Stryker Neurovascular
- FDA classification date
- May 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.
Source facts
- Action
- On 4/3/24, recall notices were mailed to Supply Chain Management/ Recall Coordinator/ Inventory Managers informing them: 1) Segregate the affected units in a secure location for return to the firm. Circulate this Recall-Removal notice internally to all interested/affected parties. 2) Maintain awareness of this communication internally until all required actions have been completed within your facility. 3) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 4) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 5) Complete and return the customer response form to nvfieldactions@stryker.com
- Initiated
- 2024-04-03
- Root cause
- Release of Material/Component prior to receiving test results
- Distribution
- US Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA, MI, MS, SC, IL.
- Recall number
- Z-1740-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206975
- FDA classified
- 2024-05-01
- Recall event number
- 94347