Skip to content

Public FDA record

Z-1741-2020: Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Pow

Recall record Z-1741-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1741-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Apr 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.

Source facts

Action
The recalling firm issued letters dated 5/8/2019 on 5/17/2019 informing the customers of the issue and that a spilled sample could lead to exposure of biohazardous material and that a software update would be made by their field service engineers at the next service call.
Initiated
2019-05-17
Root cause
Software design
Terminated
2022-03-11
Distribution
Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.
Recall number
Z-1741-2020
Recall status
Terminated
CDRH record ID
179661
FDA record created
2020-04-17
Recall event number
84918

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗