Open original FDA record ↗Recall record
- FDA identifier
- Z-1741-2020
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Apr 17, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
Source facts
- Action
- The recalling firm issued letters dated 5/8/2019 on 5/17/2019 informing the customers of the issue and that a spilled sample could lead to exposure of biohazardous material and that a software update would be made by their field service engineers at the next service call.
- Initiated
- 2019-05-17
- Root cause
- Software design
- Terminated
- 2022-03-11
- Distribution
- Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.
- Recall number
- Z-1741-2020
- Recall status
- Terminated
- CDRH record ID
- 179661
- FDA record created
- 2020-04-17
- Recall event number
- 84918