Skip to content

Public FDA record

Z-1742-2009: Stryker PainPump, 2-day infusion kit with ExFen 2.5", 120 mL infusion kit (2.08 mL/hr), sterile, Stryker

Recall record Z-1742-2009 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1742-2009
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Aug 19, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.

Source facts

Action
Stryker Instruments issued a recall letter dated June 16, 2009 and instructed Consignees to cease using and destroy the product and to return a business reply form for credit via fax to Stryker Instruments Regulatory Department, 1-866-521-2762. For further questions, contact Stryker Instruments at 1-800-800-4236 extension 3808.
Initiated
2009-06-16
Root cause
Nonconforming Material/Component
Terminated
2010-10-27
Distribution
Worldwide Distribution -- Including the United States, Argentina, Australia, Brazil, Canada, England, France, Germany, Hong Kong, Italy, Japan, Latin America, Malaysia, Mexico, Netherlands, New Zealand, Singapore, South Africa, Spain, Sweden, Switzerland and Taiwan.
Recall number
Z-1742-2009
Recall status
Terminated
CDRH record ID
83207
FDA classified
2009-08-19
Recall event number
52452

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗