Z-1742-2022: ZAP-X Radiosurgery System

Recall record Z-1742-2022 for Zap Surgical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1742-2022
Company in FDA record
Zap Surgical Systems
FDA record creation date
Sep 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

Source facts

Action
On 7/05/22 correction notices were distributed to customers who were told that revised software to address the recall issue would be supplied. Until the firm supplies the revised software follow the following steps to ensure accurate treatment delivery: 1) After using auto-align ensure initial patient setup (prior to treating the first isocenter), has rotational deviations of less than plus-minus 1.5 degrees in each dimension. 2) Prior to commencement of each isocenter beam delivery, verify that rotational deviations remain below plus-minus 1.5 degrees. - Additionally, visually verify patient alignment using image overlay before each subsequent isocenter. 3) Use "auto-align" in cases where the observed rotation deviations exceed plus-minus 1.5 degrees. This will correct the offset to within specifications. 4) Be reminded of the following warning message from IFU E0920-00004 Rev. G Page 49: "Warning: Safe and effective treatment requires accurate patient positioning. It is your responsibility to confirm the final alignment. Carefully observe the KV image and it's correction to the DRR image." The software correction will: - Correct the defect in the transitional alignment algorithm. - Revise the user interface to provide the capability to trend computed rotations in real time during treatment, to better aid the user in identifying possible alignment errors. Customers with additional questions can contact the firm at +1 (408) 933-8477, Derek@zapsurgical.com
Initiated
2022-07-05
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Recall number
Z-1742-2022
Recall status
Open, Classified
CDRH record ID
195555
FDA record created
2022-09-14
Recall event number
90716

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗