Skip to content

Public FDA record

Z-1743-2019: MRSASelect II, IVD, REF 63758 (20 plates per package). The MRSASelect is indicated for use in conjuncti

Recall record Z-1743-2019 for Bio-Rad Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1743-2019
Company in FDA record
Bio-Rad Laboratories, Inc
FDA record creation date
Jun 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abnormal coloration of agar plates and the growth of non-characteristic MRSA colonies.

Source facts

Action
The firm. Bio-Rad laboratories, Inc., issued an "URGENT PRODUCT CORRECTION" letter dated 6/7/2018 to all customers via FedEx on 6/7-8/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " As already recommended in the package insert, please ensure that exposure of agar plates to light is minimized, both before and during incubation, with the added time restriction of no more than 2 hours during storage, warming, incubation, and reading. " Do not use agar with a change in color from white to an orange-pinkish shade. " As recommended in the package insert, if in doubt, confirm the identification of colonies by coagulase test or PASTOREX STAPH-PLUS. Please feel free to contact Technical Support with any questions or concerns at 1-800-224-6723, option #2 and then #3.
Initiated
2018-06-08
Root cause
Nonconforming Material/Component
Terminated
2020-10-30
Distribution
US Nationwide Distribution to: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico.
Recall number
Z-1743-2019
Recall status
Terminated
CDRH record ID
172651
FDA record created
2019-06-06
Recall event number
82761

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗