Open original FDA record ↗Recall record
- FDA identifier
- Z-1744-2015
- Company in FDA record
- BIOMET FRANCE S.A.R.L.
- FDA classification date
- Jun 10, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The pin which maintains the knob button, may disconnect and become lost. This may result in a delay of surgery to obtain a new Optigun and detect/retrieve the pin, and may necessitate revision surgery if the pin falls into the surgical site and is not detected.
Source facts
- Action
- BIOMET sent an URGENT ADVISORY NOTICE dated May 14, 2015, to all affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to check the product to identify to determine if the pin is defective, if defective they should return it to Biomet Warsaw, following the directions on the Response Form and confirm receipt of the notice by sending back the response form. Questions related to this notice should be directed to (574) 372- 1570 Monday through Friday 8 a.m. to 5 p.m.
- Initiated
- 2015-05-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-10-27
- Distribution
- Nationwide distribution to MD only
- Recall number
- Z-1744-2015
- Recall status
- Terminated
- CDRH record ID
- 137454
- FDA classified
- 2015-06-10
- Recall event number
- 71365