Z-1744-2022: Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.

Recall record Z-1744-2022 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1744-2022
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
Sep 15, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.

Source facts

Action
The recalling firm issued letters dated August 2022 via certified mail on 8/30/2022. A black box at the beginning of the letter notates the document contains important information for the continued safe and proper use of the product. The information is to be reviewed with all members of the consignee's staff who need to be aware of the contents of the communication and to understand the implications of the communication. The letter informs the consignee what the problem is and under what circumstances it can occur and contains an embedded photograph of the Edit Prescription screen highlighting the affected fields. The hazard/harm associated with the issue is described, as well as the affected products and how to identify them. An embedded example of the About screen is contained in the letter to show where to find the affected version number. The actions to be taken by the customer/user in order to prevent risks for patients was provided. The actions include: (a) To avoid exporting an incorrect dose, do not prescribe to Percentage of Point Dose. Only prescribe plans using one of the following options provided in the letter until the software solution is provided by Philips; (b) Circulate the notice to the Head of Medical Physics in the Radiation Oncology Department or to the Department Director with specific instructions that the notice be distributed to all users of the Pinnacle3 Radiation Therapy Planning System so that they are aware of the product issue and associated hazard/harm; (c) Post the notice near the affected Pinnacle3 Radiation Therapy Planning System(s) for ease of reference; and (d) Complete and return the attached response form to Philips via email no later than 30 days from receipt of the letter. The consignee is informed that Philips is currently developing a correction which will be released in the future and once released, Philips will contact the consignee to schedule installation of the software solution to resolve the issue.
Initiated
2022-08-30
Root cause
Device Design
Distribution
Distribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Kingdom.
Recall number
Z-1744-2022
Recall status
Open, Classified
CDRH record ID
195954
FDA classified
2022-09-15
Recall event number
90819

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗