Open original FDA record ↗Recall record
- FDA identifier
- Z-1746-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- May 15, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM Definition Edge, SOMATOM Definition Flash and SOMATOM Force.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. will bring defect into Compliance noted below: 1.You will contact customers and initiate the actions to correct the failure to comply issue. 2. You will provide a software update to correct tube current calculations by the CARE Dose4D algorithm for head scans. 3.You will follow-up with consignees to verify the effectiveness of the software update. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) division office coordinator noted below are to be included. For further questions, please call (610) 448-6471.
- Initiated
- 2017-11-08
- Root cause
- Device Design
- Terminated
- 2023-05-18
- Distribution
- Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico.
- Recall number
- Z-1746-2018
- Recall status
- Terminated
- CDRH record ID
- 164181
- FDA record created
- 2018-05-15
- Recall event number
- 79295