Z-1746-2018: SOMATOM Definition Edge (Model 10590000) Computed tomography x-ray diagnostic system

Recall record Z-1746-2018 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1746-2018
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
May 15, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM Definition Edge, SOMATOM Definition Flash and SOMATOM Force.

Source facts

Action
Siemens Medical Solutions USA, Inc. will bring defect into Compliance noted below: 1.You will contact customers and initiate the actions to correct the failure to comply issue. 2. You will provide a software update to correct tube current calculations by the CARE Dose4D algorithm for head scans. 3.You will follow-up with consignees to verify the effectiveness of the software update. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) division office coordinator noted below are to be included. For further questions, please call (610) 448-6471.
Initiated
2017-11-08
Root cause
Device Design
Terminated
2023-05-18
Distribution
Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico.
Recall number
Z-1746-2018
Recall status
Terminated
CDRH record ID
164181
FDA record created
2018-05-15
Recall event number
79295

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗