Open original FDA record ↗Recall record
- FDA identifier
- Z-1746-2022
- Company in FDA record
- Qiagen Sciences LLC
- FDA record creation date
- Sep 15, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Source facts
- Action
- Customers were notified of the initial issue by email beginning Nov 3 2021, and requested to return acknowledgment of the notice. On January 26, 2022, a new version of the protocol software was released and customers were advised of the new protocol availability and acknowledge download via a follow up notice.
- Initiated
- 2021-11-21
- Root cause
- Other
- Distribution
- Domestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.
- Recall number
- Z-1746-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195665
- FDA record created
- 2022-09-15
- Recall event number