Skip to content

Public FDA record

Z-1747-2014: Boston Scientific, NovaGold High Performance Guidewire, Rx only, Sterilized using ethylene oxide, UPN M00

Recall record Z-1747-2014 for NEO METRICS, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1747-2014
Company in FDA record
NEO METRICS, INC.
FDA classification date
Jun 10, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Neo Metrics, Inc. is recalling NovaGold Guidewire because the tip may uncoil and / detach more frequently than anticipated. There have been no reports of illness or injury.

Source facts

Action
NeoMetrics sent a letter dated March 31, 2014, to Boston Scientific. On March 24, 2014, Boston Scientific made NeoMetrics aware that Boston Scientific received 3 complaints from customers that the tip of the NovaGold guidewire detached and/or became uncoiled. Because the tips are breaking and/or uncoiling at a rate that was not anticipated, it has been decided that these units should be removed from the field. Therefore, NeoMetrics does not object to Boston Scientific executing this product removal in accordance with Section 4.4 of the distribution agreement. For further questions please call (763) 559-4440.
Initiated
2014-03-31
Root cause
Nonconforming Material/Component
Terminated
2015-01-16
Distribution
Worldwide Distribution - US and the countries of Austria, Belgium, Germany, Spain, France, Great Britain, Italy, Netherlands and Sweden.
Recall number
Z-1747-2014
Recall status
Terminated
CDRH record ID
127263
FDA classified
2014-06-10
Recall event number
68279

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗