Open original FDA record ↗Recall record
- FDA identifier
- Z-1747-2014
- Company in FDA record
- NEO METRICS, INC.
- FDA classification date
- Jun 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Neo Metrics, Inc. is recalling NovaGold Guidewire because the tip may uncoil and / detach more frequently than anticipated. There have been no reports of illness or injury.
Source facts
- Action
- NeoMetrics sent a letter dated March 31, 2014, to Boston Scientific. On March 24, 2014, Boston Scientific made NeoMetrics aware that Boston Scientific received 3 complaints from customers that the tip of the NovaGold guidewire detached and/or became uncoiled. Because the tips are breaking and/or uncoiling at a rate that was not anticipated, it has been decided that these units should be removed from the field. Therefore, NeoMetrics does not object to Boston Scientific executing this product removal in accordance with Section 4.4 of the distribution agreement. For further questions please call (763) 559-4440.
- Initiated
- 2014-03-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-01-16
- Distribution
- Worldwide Distribution - US and the countries of Austria, Belgium, Germany, Spain, France, Great Britain, Italy, Netherlands and Sweden.
- Recall number
- Z-1747-2014
- Recall status
- Terminated
- CDRH record ID
- 127263
- FDA classified
- 2014-06-10
- Recall event number
- 68279