Z-1748-2017: Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29

Recall record Z-1748-2017 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1748-2017
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
Apr 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems

Source facts

Action
Siemens Healthcare issued An Urgent Medical Device Correction (UMDC) to customers in the United States on March 08, 2017 to be delivered to customers on March 09, 2017and (UFSN) was sent for distribution to customers outside the United States on March 08, 2017. These notices inform customers of the potential and mitigation action to be taken for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) assay on the ADVIA Centaur Systems. This issue affects all current and future lots of ADVIA Centaur Systems BR reagents until a solution is implemented. Fax-back forms included with letter used to confirm that the customers received the Urgent Medical Device Correction to the Customer Care Center at 312-275-7795. For questions regarding this recall call 508-668-5000. .
Initiated
2017-03-08
Root cause
Under Investigation by firm
Terminated
2020-01-22
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Australia, Brazil, Canada, Columbia, Cyprus, Japan, Spain
Recall number
Z-1748-2017
Recall status
Terminated
CDRH record ID
154128
FDA record created
2017-04-06
Recall event number
76766

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗