Skip to content

Public FDA record

Z-1749-2008: UBiT-IR300 Infrared Spectrophotometer, Software Versions 7.0, 7.01 and 7.11.

Recall record Z-1749-2008 for Otsuka America Pharmaceutical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1749-2008
Company in FDA record
Otsuka America Pharmaceutical, Inc.
FDA classification date
Sep 1, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Medical device power supply may overheat and cause smoke to be emitted from the device.

Source facts

Action
The initial distributor in the US notified consignees 03/14/08 and 03/26/08 by letter. Consignees were provided two different letters depending on the hours of use of the spectrophotometer. The notifications advised of the firm's preventive service and upgrade program for the instrument's power supply, in which consignees would receive a temporary replacement instrument while the service is completed. Consignees with spectrophotometers having <15000 light source hours were notified similarly, and were provided filters and instructions for cleaning and replacing the filters to protect the power supply from unnecessary accumulation of debris. Contact firm at 1-888-MERETEK for assistance.
Initiated
2008-01-18
Root cause
Nonconforming Material/Component
Terminated
2011-02-18
Distribution
The affected medical devices were distributed to medical facilities and physician offices nationwide and to distributors in Mexico.
Recall number
Z-1749-2008
Recall status
Terminated
CDRH record ID
69717
FDA classified
2008-09-01
Recall event number
47536

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.