Skip to content

Public FDA record

Z-1750-2015: Astral 100 ( Model No. 27001) & Astral 150 (Model No. 27003) ventilator.

Recall record Z-1750-2015 for ResMed Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1750-2015
Company in FDA record
ResMed Ltd.
FDA classification date
Jun 10, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can lead to hazards of insufficient ventilation and insufficient expiratory pressure which can result in major severity harm.

Source facts

Action
A customer notification letter dated 5/5/15 will be sent to customers that the Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. The letter provides the customers with the hazards involved and the actions to be taken. Customers are instructed to ensure alarms are appropriately set up for ventilator dependent patients. Customers with questions are instructed to contact Larissa D'Andrea at (858) 836-6837 or email larissa.d'andrea@resmed.com.
Initiated
2015-05-05
Root cause
Use error
Terminated
2016-04-07
Distribution
Worldwide distribution. US nationwide, Uruguay and Canada.
Recall number
Z-1750-2015
Recall status
Terminated
CDRH record ID
136099
FDA classified
2015-06-10
Recall event number
71100

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗