Open original FDA record ↗Recall record
- FDA identifier
- Z-1751-2022
- Company in FDA record
- Alcor Scientific, Inc.
- FDA record creation date
- Sep 16, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
Source facts
- Action
- Alcor Scientific notified consignees via Emails on 21 July 2022 that included an Advisory Notice (AN-001). Letter states reason for recall, health risk and action to take: " Remove from your inventory any Lot 4/22 Primary Pump Tubing Spare Part Kits (112-13-002). Destroy this tubing. " If the tubing has been installed on any iSED ESR Analyzers, the tubing must be immediately replaced with new Primary Pump Tubing (112-13-002) from any other lot and Alcor must be notified of the locations where that tubing was installed. " Complete this form and return via email to techservice@alcorscientific.com within 3 days. ALCOR will replace all Primary Pump Tubing from Lot 4/22 that was shipped to your facility.
- Initiated
- 2022-07-20
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
- Recall number
- Z-1751-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195684
- FDA record created
- 2022-09-16
- Recall event number
- 90762