Z-1751-2022: iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the a

Recall record Z-1751-2022 for Alcor Scientific, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1751-2022
Company in FDA record
Alcor Scientific, Inc.
FDA record creation date
Sep 16, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube

Source facts

Action
Alcor Scientific notified consignees via Emails on 21 July 2022 that included an Advisory Notice (AN-001). Letter states reason for recall, health risk and action to take: " Remove from your inventory any Lot 4/22 Primary Pump Tubing Spare Part Kits (112-13-002). Destroy this tubing. " If the tubing has been installed on any iSED ESR Analyzers, the tubing must be immediately replaced with new Primary Pump Tubing (112-13-002) from any other lot and Alcor must be notified of the locations where that tubing was installed. " Complete this form and return via email to techservice@alcorscientific.com within 3 days. ALCOR will replace all Primary Pump Tubing from Lot 4/22 that was shipped to your facility.
Initiated
2022-07-20
Root cause
Nonconforming Material/Component
Distribution
US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
Recall number
Z-1751-2022
Recall status
Open, Classified
CDRH record ID
195684
FDA record created
2022-09-16
Recall event number
90762

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗