Skip to content

Public FDA record

Z-1752-2014: Stryker 4.0mm Round Fluted Bur Sterile Product Usage: The devices are intended to be used to cut bone an

Recall record Z-1752-2014 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1752-2014
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Jun 11, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There may be missing notches, partial notches or no notches at all in some Cutting Accessories that are used with the Elite" Attachments.

Source facts

Action
Stryker Instruments sent an URGENT MEDICAL DEVICE RECALL notification letter dated May 22, 2014, to the consignees. Recall notification included affected product description and identification, reason for recall, risk to health, and actions to be taken by the customers. Customers will be instructed to review the notification, quarantine unused product and to return a signed Business Reply Form. They will be advised that a Stryker Representative will be sent to their facility to inspect any affected product on hand. Once product is inspected by the Stryker Representative, they can return conforming product back to stock, for normal use. Non-conforming product will be removed from the facility by a Stryker Representative following the inspection process. The Stryker Representative will send the product to the manufacturer for destruction. For questions regarding this recall please contact Stryker Instruments: Kara Spath 269-389-4518 kara.spath@stryker.com
Initiated
2014-05-22
Root cause
Nonconforming Material/Component
Terminated
2015-05-06
Distribution
Worldwide Distribution - US (nationwide)
Recall number
Z-1752-2014
Recall status
Terminated
CDRH record ID
127411
FDA classified
2014-06-11
Recall event number
68331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗