A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.
Source facts
Action
The customers received an "Urgent Device Correction" notification letter on May 12, 2013 explaining the product issue and providing recommended actions. Customers were contacted to set up a service call to test and recalibrate equipment as necessary. Questions were directed to 800-778-7898.
Initiated
2013-04-25
Root cause
Employee error
Terminated
2013-07-26
Distribution
Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.
Recall number
Z-1753-2013
Recall status
Terminated
CDRH record ID
118118
FDA classified
2013-07-18
Recall event number
65176