RaySert PLUS RSP01 LOT K030 labeled with an expiry of 2024-01 contained an inner tray that was labeled with an expiry of 2015. In addition to having an expiry of 2015 the LOT number on the inner tray is identified as V001.
Source facts
Action
Rayner lntraocular notified customers on 03/17/2020 via "Urgent Medical Device recall" letter. The recall letter identified the affected product and lot number. The customers were asked to immediately quarantine and return the product.
Initiated
2020-03-17
Root cause
Labeling mix-ups
Distribution
US nationwide distribution in the states of OH, FL, GA, MD, SC, NJ, PR, IN, ID and MN.
Recall number
Z-1753-2020
Recall status
Open, Classified
CDRH record ID
180399
FDA record created
2020-04-21
Recall event number
85270