Skip to content

Public FDA record

Z-1756-2024: The Oncology Essentials Package of the Spectral CT imaging system.

Recall record Z-1756-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1756-2024
Company in FDA record
Philips North America Llc
FDA classification date
May 2, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.

Source facts

Action
An URGENT Medical Device Correction Letter, dated 3/27/24, was mailed to consignees to notify them of this recall. Consignees are instructed to identify if a third-party laser was installed on their system and if the laser line is reflected, duplicated, or thickened. If a third-party laser is installed, before treatment of a patient, reposition or re-image as necessary to verify proper placement of the localization laser marks. Systems may continue to be used in accordance with the intended use and provided use recommendations. Consignees are to complete and return the provided response form by email to CTNM.QARA@philips.com. The provided recall notification is to be disseminated to all users of affected systems. Philips will contact consignees to schedule a time for a Field Service Engineer to visit and install hardware to resolve the issue. Questions are to be directed to the Customer Care Solutions Center at 1-800-722-9377 from 8:00 AM to 8:00 PM, Monday through Friday.
Initiated
2024-03-29
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
Recall number
Z-1756-2024
Recall status
Open, Classified
CDRH record ID
206995
FDA classified
2024-05-02
Recall event number
94352

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗