Open original FDA record ↗Recall record
- FDA identifier
- Z-1757-2010
- Company in FDA record
- Cardiac Science Corporation
- FDA classification date
- Jun 4, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
Source facts
- Action
- On April 3, 2009 the recalling firms notified consignees by certified letter to domestic customers and via FedEx to international customers with delivery confirmation signature required. Customers were advised to remove treadmill from service until it is tested by a Field Service Engineer. For customers with questions or those that have not already scheduled a service call with Cardiac Science Field Service Engineers, please contact the firm at 1.800.426.0337, option 1.
- Initiated
- 2009-03-30
- Root cause
- Other
- Terminated
- 2012-01-10
- Distribution
- Distributed domestically (including Puerto Rico) to 20 end-user accounts and 2 distributor accounts; Distributed internationally to: Canada, Mexico, Portugal, and Saudi Arabia.
- Recall number
- Z-1757-2010
- Recall status
- Terminated
- CDRH record ID
- 81323
- FDA classified
- 2010-06-04
- Recall event number
- 51812