Open original FDA record ↗Recall record
- FDA identifier
- Z-1758-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Jun 4, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated after Beckman Coulter confirmed reports of potential dilution of reagent vials loaded onto the Preppies or PrepPlus 2 systems. Affected systems have a syringe pump with part number A46748 (see illustrations below) which cause the probe to prematurely dispense Isoflow prior to arriving at the wash station. This condition only occurs with worklist panels which include both
Source facts
- Action
- A Product Corrective Action (PCA) letter with attached PCA Response Form was sent on the week of January 25, 2010 to all customers who purchased the pipetting and diluting system: COULTER PrepPlus, or COULTER PrepPlus 2. The letter provides the customers with an explanation of the problem identified and an instruction on proper placing antibody reagent when using a worklist panel which involves both blood specimens and antibody reagents. Customers were also instructed to complete and return the PCA Response Form in addition to sharing the information contained in the PCA with their laboratory staff. If consignees had any questions they were told to contact their local Beckman Coulter Representative. Questions can be directed to the company's Customer Service office at (800) 526-7694.
- Initiated
- 2010-01-21
- Root cause
- Other
- Terminated
- 2012-04-04
- Distribution
- Domestically and Canada.
- Recall number
- Z-1758-2010
- Recall status
- Terminated
- CDRH record ID
- 90202
- FDA classified
- 2010-06-04
- Recall event number
- 55166