Skip to content

Public FDA record

Z-1758-2018: Retractor f/Sciatic Nerve; Part number: 03.100.013; UDI: 10886982070623

Recall record Z-1758-2018 for Synthes (USA) Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1758-2018
Company in FDA record
Synthes (USA) Products LLC
FDA record creation date
May 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for microspores to form on the hollow handle of the Sciatic Nerve Retractor. The pores may increase in size, allowing fluid to enter the hollow handle.

Source facts

Action
On April 2, 2018 Synthes (USA) LLC mailed an Urgent Product Recall notification letter to affected customers. Customers were instructed to: Review current inventory to identify and quarantine all affected products; Complete the Verification section of the Return Receipt form; Return affected product; Forward the notification to anyone in your facility that needs to be informed; If affected products have been forwarded to another facility, contact that facility to arrange return; Keep a copy of the notification. Customers with questions may call (610)719-5450 or contact your Depuy Synthese Sales Consultant.
Initiated
2018-04-02
Root cause
Under Investigation by firm
Terminated
2019-10-16
Distribution
US Nationwide and Canada
Recall number
Z-1758-2018
Recall status
Terminated
CDRH record ID
163734
FDA record created
2018-05-09
Recall event number
79850

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.