Z-1758-2020: ClearCanvas RIS/PACS

Recall record Z-1758-2020 for Synaptive Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1758-2020
Company in FDA record
Synaptive Medical Inc
FDA record creation date
Apr 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect negatively affects the quality of images taken during diagnostic imaging (mammograms) which may result in misdiagnosis.

Source facts

Action
The firm initiated this field corrective action by letter on 02/26/2020. The consignees are not required to discontinue using the device nor return the device to Synaptive Medical, a workaround is to be followed as indicated in the customer notification letter. The actions to be taken by consignees with the affected product are as follows: 1. Acknowledge receipt of the notification using the acknowledgment and receipt form that the firm has provided. 2. a) If ClearCanvas Workstation is NOT used for viewing images on removeable media such as CDs, and this issue is encountered, configure the PACS or DICOM device that is sending images to ClearCanvas Workstation to send uncompressed images, or images compressed with a compression algorithm other than JPEG 2000. b.) If ClearCanvas Workstation IS used for viewing images on removeable media such as CDs OR if it is not possible to change the sending device to send uncompressed images, contact Synaptives Product Support Recall Line at 1.866.853.8001 to request assistance to correct the device. Note that if the software is subsequently re-installed, the device will need to be corrected again. Contact Synaptive to repeat the correction.
Initiated
2020-02-26
Root cause
Under Investigation by firm
Terminated
2024-04-10
Distribution
US
Recall number
Z-1758-2020
Recall status
Terminated
CDRH record ID
180296
FDA record created
2020-04-22
Recall event number
85216

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗