Skip to content

Public FDA record

Z-1759-2013: AlboGraft Knitted with Collagen, Bifurcate 20x10Catalog Number: AMC2010 AlboGraft Vascular Grafts are m

Recall record Z-1759-2013 for LeMaitre Vascular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1759-2013
Company in FDA record
LeMaitre Vascular, Inc.
FDA classification date
Jul 22, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Blood blushing/leaking from the surface of the graft after implantation

Source facts

Action
LeMaitre Vascular, Inc.sent an Urgent Field Safety Notice daed June 19, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and return the affected device to MeMaitre Vascular which they will replace with a new unit. Customers with questions were instructed to call 781-221-2266, ext. 183. For questions regarding this recall call 781-425-1670, ext 108.
Initiated
2013-06-17
Root cause
Process control
Terminated
2016-08-15
Distribution
Worldwide distribution: USA including PA and countries of: BRAZIL, BULGARIA, FRANCE, GEORGIA, GERMANY, GREECE, ITALY, LEBANON, NETHERLAND, POLAND, SPAIN, and TURKEY.
Recall number
Z-1759-2013
Recall status
Terminated
CDRH record ID
119548
FDA classified
2013-07-22
Recall event number
65553

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗