Open original FDA record ↗Recall record
- FDA identifier
- Z-1759-2018
- Company in FDA record
- Synthes (USA) Products LLC
- FDA record creation date
- May 9, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for microspores to form on the hollow handle of the Sciatic Nerve Retractor. The pores may increase in size, allowing fluid to enter the hollow handle.
Source facts
- Action
- On April 2, 2018 Synthes (USA) LLC mailed an Urgent Product Recall notification letter to affected customers. Customers were instructed to: Review current inventory to identify and quarantine all affected products; Complete the Verification section of the Return Receipt form; Return affected product; Forward the notification to anyone in your facility that needs to be informed; If affected products have been forwarded to another facility, contact that facility to arrange return; Keep a copy of the notification. Customers with questions may call (610)719-5450 or contact your Depuy Synthese Sales Consultant.
- Initiated
- 2018-04-02
- Root cause
- Under Investigation by firm
- Terminated
- 2019-10-16
- Distribution
- US Nationwide and Canada
- Recall number
- Z-1759-2018
- Recall status
- Terminated
- CDRH record ID
- 163736
- FDA record created
- 2018-05-09
- Recall event number
- 79850