Product found to not comply to Class 1 Laser safety requirement.
Source facts
Action
Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.
Initiated
2022-02-28
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-1762-2022
Recall status
Open, Classified
CDRH record ID
195985
FDA classified
2022-09-26
Recall event number
90835
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.