Skip to content

Public FDA record

Z-1763-2020: Olympus Pleuravideoscope, Model LTF-240

Recall record Z-1763-2020 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1763-2020
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Apr 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was no data to validate that the identified endo-therapy devices were compatible with certain pleuravideoscope models. Updated Operations Manuals are being provided.

Source facts

Action
An Urgent Medical Device Correction notification letter dated 3/23/20 was sent to customers. Actions to be taken by the user: Our records indicate that your facility has purchased one or more of the above-referenced endoscopes. OLYMPUS requires that you take the following actions: a) Inspect your inventory of operation manuals for the referenced Olympus models and discard the existing ones. b) Exchange them with the updated, enclosed Operation Manuals having an updated system chart indicating compatible endoscope accessories. c) Notify related personnel at your institution that they should stop the combination use of the LTF-160 and LTF-240 and the subject Endo-Therapy accessories. d) Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0066 and provide your contact information as indicated in the portal. e) If you have further distributed this product, identify your customers, forward them this notification including the attachments, and appropriately document your notification process. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at laura.storms@olympus.com.
Initiated
2020-03-23
Root cause
Under Investigation by firm
Terminated
2021-11-30
Distribution
US Nationwide
Recall number
Z-1763-2020
Recall status
Terminated
CDRH record ID
180281
FDA record created
2020-04-22
Recall event number
85208

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗