Z-1763-2022: TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only M

Recall record Z-1763-2022 for Cardiac Assist, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1763-2022
Company in FDA record
Cardiac Assist, Inc
FDA classification date
Sep 23, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.

Source facts

Action
On July 21, 2022, the recalling firm notified impacted customers through mailed letters. Customers were instructed to use Operations Manual revision 15 and to discard all previous revisions of the Operations Manuals.
Initiated
2022-07-21
Root cause
Under Investigation by firm
Terminated
2023-08-08
Distribution
US Nationwide distribution.
Recall number
Z-1763-2022
Recall status
Terminated
CDRH record ID
195659
FDA classified
2022-09-23
Recall event number
90755

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗