Open original FDA record ↗Recall record
- FDA identifier
- Z-1763-2025
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- May 14, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
Source facts
- Action
- Firm began notifying consignees on May 2, 2025 via hand delivery of letters titled "Urgent Medical Device Correction." Issue 1: User receives an error if they attempt to define a Percentage Dose Change Increase limit between 60% and 99%. Customers were instructed to ensure all users or potential users of the software are made aware of the notification and proposed mitigations: Use Percentage Dose Change Increase values between 20% and 59% or between 100% and 500%. Pharmacists will not be able to use values from 60% to 99%. Issue 2: It is possible for an unauthorized user to bypass the reauthentication steps required to approve a drug library or delete a medication ruleset. Customers were instructed to ensure all users or potential users are made aware of the notification and proposed mitigation: Communicate with all users the need to log off after finishing a session before leaving the workstation and following cybersecurity best practices. ICU Medical is developing a LifeShield software update to correct both issues.
- Initiated
- 2025-05-02
- Root cause
- Software Design Change
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.
- Recall number
- Z-1763-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213581
- FDA classified
- 2025-05-14
- Recall event number
- 96771