Skip to content

Public FDA record

Z-1763-2025: LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2

Recall record Z-1763-2025 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1763-2025
Company in FDA record
ICU Medical, Inc.
FDA classification date
May 14, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

Source facts

Action
Firm began notifying consignees on May 2, 2025 via hand delivery of letters titled "Urgent Medical Device Correction." Issue 1: User receives an error if they attempt to define a Percentage Dose Change Increase limit between 60% and 99%. Customers were instructed to ensure all users or potential users of the software are made aware of the notification and proposed mitigations: Use Percentage Dose Change Increase values between 20% and 59% or between 100% and 500%. Pharmacists will not be able to use values from 60% to 99%. Issue 2: It is possible for an unauthorized user to bypass the reauthentication steps required to approve a drug library or delete a medication ruleset. Customers were instructed to ensure all users or potential users are made aware of the notification and proposed mitigation: Communicate with all users the need to log off after finishing a session before leaving the workstation and following cybersecurity best practices. ICU Medical is developing a LifeShield software update to correct both issues.
Initiated
2025-05-02
Root cause
Software Design Change
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.
Recall number
Z-1763-2025
Recall status
Open, Classified
CDRH record ID
213581
FDA classified
2025-05-14
Recall event number
96771

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗