Open original FDA record ↗Recall record
- FDA identifier
- Z-1764-2008
- Company in FDA record
- Focus Diagnostics Inc
- FDA classification date
- Aug 17, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate interpretation of a sample as negative.
Source facts
- Action
- Focus notified all clients which received the recalled lots starting on July 6, 2007 via an Important Medical Device Information letter by fax or email. The letter explains how to identify suspect results and the steps to be taken if suspect results are obtained. If customers find the interpretation of results with the indicated kit lots to be problematic, customers have the option of having their kits replaced at no charge within 90 days of receiving the notification.
- Initiated
- 2007-07-06
- Root cause
- Process design
- Terminated
- 2009-06-25
- Distribution
- Worldwide Distribution --- USA including states of OH, CA, CT, IL, NY, and countries of Barbados, Canada, Germany, Israel, Slovakia, France, and England.
- Recall number
- Z-1764-2008
- Recall status
- Terminated
- CDRH record ID
- 69534
- FDA classified
- 2008-08-17
- Recall event number
- 47459