Open original FDA record ↗Recall record
- FDA identifier
- Z-1764-2010
- Company in FDA record
- Angiodynamics, Inc.
- FDA classification date
- Jun 4, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product has the potential for a cross-lumen leak. One of the two leak testers that are used to perform the leak test on the DuraMax Catheters was found to be faulty. The tester was not able to detect cross-lumen leaks. Both testers were used during the manufacturing 100% leak test.
Source facts
- Action
- Customers were sent URGENT MEDICAL DEVICE RECALL LETTERS (dated 1/07/10) via Certified Mail, Return Receipt Requested. Customers were instructed to return all recalled product. A return authorization number was assigned to each customer in the recall notification letter and a UPS account was provided for return freight. A reply form was provided in the recall notification letter. Questions were directed to the company sales representatives or to Customer Service at 800-772-6446.
- Initiated
- 2010-01-08
- Root cause
- Other
- Terminated
- 2012-07-25
- Distribution
- Nationwide and St. Thomas, VI
- Recall number
- Z-1764-2010
- Recall status
- Terminated
- CDRH record ID
- 91605
- FDA classified
- 2010-06-04
- Recall event number
- 55701