Open original FDA record ↗Recall record
- FDA identifier
- Z-1765-2008
- Company in FDA record
- Kyphon Inc
- FDA classification date
- Sep 29, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Physician instructions revised due to product breakage: During the implantation of the X-STOP IPD Device, the product has the potential to cause damage to and/or breakage of the X-STOP IPD Device's universal wing assembly. This includes the potential for breakage of the wing screw.
Source facts
- Action
- On January 30, Kyphon sent a letter to all consignees, and a teleconference was held with sales management informing them of the recall and the need to contact consignees. The product labeling including the Physician Training and Physician's Guide have been updated to reflect the potential risk of breakage.
- Initiated
- 2008-01-30
- Root cause
- Labeling design
- Terminated
- 2009-07-14
- Distribution
- Worldwide distribution: USA, UK, Netherlands, Canada, Australia, Denmark, Germany, Greece, Hong Kong, Israel, Italy, Philippines, Portugal, Singapore, South Africa, Spain Switzerland, Turkey and UAE.
- Recall number
- Z-1765-2008
- Recall status
- Terminated
- CDRH record ID
- 69581
- FDA classified
- 2008-09-29
- Recall event number
- 47490