Open original FDA record ↗Recall record
- FDA identifier
- Z-1765-2017
- Company in FDA record
- ORTHO-CLINICAL DIAGNOSTICS
- FDA record creation date
- Apr 8, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.
Source facts
- Action
- On 3/7/2017, IMPORTANT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2017-048) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers and distributors who were shipped VITROS NT-proBNP Reagent Packs, Lots 1570, 1580 and/or 1590 to inform them of the issue and provided further instructions.
- Initiated
- 2017-03-07
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-06-14
- Distribution
- Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.
- Recall number
- Z-1765-2017
- Recall status
- Terminated
- CDRH record ID
- 153781
- FDA record created
- 2017-04-08
- Recall event number
- 76665