Open original FDA record ↗Recall record
- FDA identifier
- Z-1766-2026
- Company in FDA record
- Oxoid Australia Pty Limited
- FDA classification date
- Apr 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells.
Source facts
- Action
- Thermo Fisher Scientific issued an Urgent: Medical Device Recall notice to its consignees on 3/3/2036 via USPS First Class Mail. The notice explained problem, risk to health, requested the consignee destroy any remaining inventory of the affected lot and responding by completing the Acknowledgement form and returning it via email to MBD.vigilance@thermofisher.com in order to receive replacement/credit to be processed keeping the notification on file. The notice further requests it be passed on to all who need to be aware within the consignee's organization or to any organization where the potentially affected products have been transferred. For questions, please contact our Technical Services Department at 800-255-6730 (US) or email microbiology.ts.us@thermofisher.com.
- Initiated
- 2026-03-03
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of GA and CA.
- Recall number
- Z-1766-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219024
- FDA classified
- 2026-04-02
- Recall event number
- 98539