Open original FDA record ↗Recall record
- FDA identifier
- Z-1767-2020
- Company in FDA record
- Philips North America, LLC
- FDA classification date
- Apr 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
Source facts
- Action
- Urgent - Medical Device Recall notification letters dated 3/5/20 were sent to customers. " Confirm whether PIC iX Release C.02.xx and C.03.01 is used anywhere in your facility, and if so: o Identify Philips TRx4841A and TRx4851A Telemetry Transceivers in your facility by using the instructions provide in the HOW TO IDENTIFY AFFECTED PRODUCTS section above. o Immediately discontinue use of the Philips TRx4841A and TRx4851A Telemetry Transceiver. Philips support for these devices ended in 2017. " Complete, sign and return the reply form provided on the last page of this letter, as directed on the form. Philips will contact you should further action be necessary. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800-722-9377
- Initiated
- 2020-03-05
- Root cause
- Software design
- Terminated
- 2021-02-11
- Distribution
- The products were distributed US nationwide. Global distribution.
- Recall number
- Z-1767-2020
- Recall status
- Terminated
- CDRH record ID
- 180352
- FDA classified
- 2020-04-23
- Recall event number
- 85241