Open original FDA record ↗Recall record
- FDA identifier
- Z-1768-2008
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Sep 4, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Images may potentially shift, causing an incorrect alignment of the patient, which may result in a dose to the wrong location. It was discovered through the complaint handling process that a potential image shift issue can occur after using the interactive shift tool and filtering tool due to incorrect consideration of the flat panel alignment.
Source facts
- Action
- A Customer Safety Advisory Notice was issued on September 19, 2007--All affected consignees were notified certified mail (includes update instruction TH029/07/S). Customers were request to maintain a copy of the safety advisory notice with the COHERENCE Therapist R2.1/Primeview 3i manuals. A copy of the firm's corrections & removal report will be forwarded to Quality Management of Siemens Regional units responsible for reporting to local country authorities and a copy of the report is also forwarded to the firm's distributor.
- Initiated
- 2007-09-19
- Root cause
- Software design
- Terminated
- 2010-12-21
- Distribution
- Class II Recall - Worldwide Distribution --- including USA, Algeria, Argentina, Australia, Canada, Colombia, Croatia, Egypt, France, Germany, Hungary, India, Ireland, Italy, Japan, New Zealand, P.R. China, Poland, Russian Fed., Saudi Arabia, South Africa, Spain, Turkey.
- Recall number
- Z-1768-2008
- Recall status
- Terminated
- CDRH record ID
- 70071
- FDA classified
- 2008-09-04
- Recall event number
- 47756