Z-1769-2019: Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog

Recall record Z-1769-2019 for Terumo Cardiovascular Systems Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1769-2019
Company in FDA record
Terumo Cardiovascular Systems Corporation
FDA record creation date
Jun 11, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.

Source facts

Action
On October 17, 2018, the firm notified customers of the recall via Urgent Medical Device Recall letter. The letter alerted customers that the devices were assembled with stepper motors that did not meet supplier specifications. Customers were informed that in the event of a failure, the gas system flow and FiO2 control sliders on the Central Control Monitor would be disabled and the gas system could only be controlled by the local control knobs. Customers were asked to take the following actions: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to schedule the field correction activities. Note: Terumo CVS recommends that users continue using Terumo System 1 while waiting for this correction. If you have any questions, please contact Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET Recall Fax: 1.734.741.6149
Initiated
2018-10-17
Root cause
Process control
Terminated
2020-04-30
Distribution
Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.
Recall number
Z-1769-2019
Recall status
Terminated
CDRH record ID
172970
FDA record created
2019-06-11
Recall event number
82878

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗