Z-1770-2019: Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294, 10015414, 10933805, 1117144

Recall record Z-1770-2019 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1770-2019
Company in FDA record
Becton Dickinson & Company
FDA classification date
Jul 11, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.

Source facts

Action
The firm, BD, sent "Urgent Medical Device Recall" letters dated 5/6/19 to its customers. The letters described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Immediately review your inventory for the specific model codes and lot numbers listed in the Affected Product Table. Destroy all product subject to the recall based on the affected product table -following your institutions process for destruction. 2. BD is instructing customers to discard all model codes and lot numbers with an expiration date between and including 05/2019 and 08/2020 in addition to the Affected Product List in this communication. 3. Share the recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 4. Until all affected product has been discarded and is no longer in use, please follow the instruction below: During infusion of critical medications, check the medication bag more frequently than normal to ensure that the remaining volume corresponds to the expected delivery time. 5. As always, review and follow the Instructions for Use provided with the product and the Alaris System, which includes the following: After the infusion set is loaded into the pump with the door closed, open the roller clamp and verify that no fluid is dripping through the drip chamber. If drips are observed, do not initiate infusion with the Pump Module unit and infusion set. Ensure that the clamp on the Alaris Pump Module Infusion Set is closed whenever infusion is not intended. 6. Any case of over-infusion or other serious or unexpected medical device incidents in patients receiving infusions with these Infusion Sets should be reported to Becton Dickinson. When reporting, please identify the model code and lot number to aid in the investigation process. 7. Complete the attached Customer Response Form and return to the BD contact noted on the form whether you have any of the impacted
Initiated
2019-05-06
Root cause
Nonconforming Material/Component
Terminated
2021-07-19
Distribution
Worldwide Distribution; US (nationwide distribution) to following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, ID, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY, and Puerto Rico; and countries of: Australia, Belgium, Canada, Chile, Malaysia, Mexico, Panama, Singapore and Taiwan.
Recall number
Z-1770-2019
Recall status
Terminated
CDRH record ID
172627
FDA classified
2019-07-11
Recall event number
82752

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗