Z-1772-2021: MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN; PUMP MMT-1740KX 670G V6.3 CLIN 0

Recall record Z-1772-2021 for Medtronic Minimed. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1772-2021
Company in FDA record
Medtronic Minimed
FDA record creation date
Jun 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.

Source facts

Action
On 03/15/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via regionally approved methods (e.g. email, registered mail, courier, hand delivery) informing customers that their insulin pump with software version 6.5 that could present pump errors after delivering a large bolus under certain conditions. These errors may occur if All the following conditions are met: 1-The bolus delivery speed is programmed to Quick in the pump settings (default is Standard). 2-The SmartGuard" feature is in use. 3-The pump needs to be on the bolus delivery screen when an auto correction bolus is triggered. Auto corrections are triggered when sensor glucose (SG) is running high and Active Insulin is low. 4-The bolus amount programmed to be delivered is greater than 17.1U. Note: This can be a single bolus greater than 17.1U or a combination of boluses totaling greater than 17.1U. If all the above conditions are met, within 2 minutes of bolus delivery completion, the pump initiates Pump error 53, followed by Pump error 23 alarm. Upon clearing the pump errors, the pump resets and indicates Active Insulin has been cleared. The pump then guides the user to resume operation in Manual Mode. The SmartGuard" status screen indicates the warm-up period has started. After approximately 5 hours, the SmartGuard" feature will be available. Since Active Insulin will display 0.0 units in the pump after experiencing the pump errors above, if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery, which may result in low blood sugar (hypoglycemia) or severe hypoglycemia. In rare cases, severe hypoglycemia, if left untreated, may lead to a life-threatening situation. As of January 21, 2021, Medtronic has received no reports of serious patient harm or serious injury related to this issue. Customers are instructed to: If their pump screen indicates that Active Insulin has cleare
Initiated
2021-03-15
Root cause
Software design
Distribution
U.S.(Clinical): AR, CA, CO, CT, FL, GA, ID, MI, MN, TN, TX, and WA O.U.S.: Austria, Bahrain, Belgium, Canary Islands, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Liechtenstein, Luxembourg, Malta, Martinique, Netherlands, New Caledonia, Norway, Poland, Qatar, Reunion, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom
Recall number
Z-1772-2021
Recall status
Open, Classified
CDRH record ID
187401
FDA record created
2021-06-01
Recall event number
87735

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗