Z-1772-2026: MicroScan Neg MIC 3J REF C54814

Recall record Z-1772-2026 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1772-2026
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Apr 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

Source facts

Action
On 07/15/2024, firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customer inform them that customers have reported multiple patient samples with skipped wells for Ceftolozane/Tazobactam (C/T) when tested with MicroScan Neg MIC 3J Lot 2025-08-23. The C/T 2/4 dilution well exhibited no growth while growth was observed in the C/T 4/4 well. Customers are instructed to: "Discontinue use of this lot and discard per your laboratory guidelines. "Consider reviewing the C/T antimicrobial susceptibility test results from this lot if Enterobacterales reported as resistant and P. aeruginosa reported as intermediate or if an increasing trend in resistance has been observed for C/T in your laboratory. "The laboratory should retain the inventory of other MicroScan Neg MIC 3J lots as they are not impacted by this issue. any questions regarding this notice, please contact us: "From our website: http://www.beckmancoulter.com "By phone: please contact: Customer call center TEL :0120-566-730
Initiated
2025-06-25
Root cause
Process control
Distribution
International distribution in the country of Japan.
Recall number
Z-1772-2026
Recall status
Open, Classified
CDRH record ID
219075
FDA classified
2026-04-03
Recall event number
98553

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗