Open original FDA record ↗Recall record
- FDA identifier
- Z-1773-2013
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Jul 19, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The LIFEPAK 500 Automated External Defibrillator (AED) could be susceptible to malfunction due to an assembly error. If affected, the malfunction would be the result of internal shorting on a printed circuit board assembly and could render the device inoperable.
Source facts
- Action
- The firm, Physio-Control Inc., contacted their consignees by telephone or email starting on June 13, 2013 and sent and/or delivered an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 2013 to its customers. The letter described the product, problem and actions to be taken. A Physio-Control representative will be contacting the customers by direct visit, telephone, or email you to coordinate the replacement of their device(s). Customers with questions regarding this notification, please call Physio's Technical Support at 1-800-442-1142, 6:00 a.m. to 4:00. pm. (Pacific), Monday-Friday.
- Initiated
- 2013-06-12
- Root cause
- Employee error
- Terminated
- 2013-09-19
- Distribution
- Worldwide distribution: US (nationwide) including states of: IL, MA, NC, ND, NY, PA, RI, TX, WA, and WI; and country of: Canada.
- Recall number
- Z-1773-2013
- Recall status
- Terminated
- CDRH record ID
- 119584
- FDA classified
- 2013-07-19
- Recall event number
- 65565