Open original FDA record ↗Recall record
- FDA identifier
- Z-1773-2017
- Company in FDA record
- Draegar Medical Systems, Inc.
- FDA record creation date
- Apr 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
Source facts
- Action
- On March 28, 2017, US consignees were sent an Urgent Medical Recall letter and Customer Reply and Order card. Delta family monitors running software version 10.0 in facilities that also have at least one Scio module will be downgraded to software version VF9.1 as a temporary solution Free of Charge. Draeger Medical Systems, Inc. is developing a new software version to resolve the issue (VF10.1). Once available, all Delta family monitors that were running software version VF10.0 will be upgraded with VF10.1 Free of Charge. For all other countries, the responsible local Draeger Sales and Service will mail the letter and card.
- Initiated
- 2017-03-28
- Root cause
- Under Investigation by firm
- Terminated
- 2018-09-17
- Distribution
- Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
- Recall number
- Z-1773-2017
- Recall status
- Terminated
- CDRH record ID
- 154430
- FDA record created
- 2017-04-10
- Recall event number
- 76862