Z-1773-2021: Change Healthcare Enterprise Viewer - Product Usage: intended to be used with off-the-shelf hardware for

Recall record Z-1773-2021 for CHANGE HEALTHCARE CANADA COMPANY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1773-2021
Company in FDA record
CHANGE HEALTHCARE CANADA COMPANY
FDA record creation date
Jun 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect was identified where the Image Styles defined by presentation states from CHRS are not displayed in CHEV.

Source facts

Action
On 04/19/21 the firm sent a letter to its consignees with the following instructions: Immediate actions to take to reduce the risk to patient Prior to treating patients, always confirm the correct patient anatomy by reviewing the laterality markers applied during image acquisition. When the issue is detected, the clinician is able to manually adjust the Image Styles in CHEV to accommodate their viewing requirements and/or orient images to standard anatomical position. Product updates that will address the issue Change Healthcare is developing a software update to correct image display of orientation, image inversion and window level information from Image Styles. Recommended actions to take on potentially affected data We strongly encourage customers receiving this Urgent Field Safety Notice to immediately contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice; and, 2) Arrange the installation of software update. Change Healthcare Customer Support: 1-800-663-2533 Menu Option 3 (CHEV)
Initiated
2021-04-19
Root cause
Software design
Distribution
US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IA, IN, IL, KY, NE, NY, MA, MI, MS, OH, PA, TX,TN, WA, WI, WY.
Recall number
Z-1773-2021
Recall status
Open, Classified
CDRH record ID
187324
FDA record created
2021-06-02
Recall event number
87891

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗