Z-1773-2026: XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.

Recall record Z-1773-2026 for XTANT Medical Holdings, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1773-2026
Company in FDA record
XTANT Medical Holdings, Inc
FDA classification date
Apr 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.

Source facts

Action
Xtant Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignee on 2/18/2026 via FedEx. The notice explained the reason for the correction, risk to health, and requested the following: 1. Review all devices from Lot 4375-01, including devices in loose stock and devices staged in implant trays or caddies, and verify device identification using the laser-etched markings on each implant to confirm that devices are stored or staged in the appropriate location consistent with their size and configuration. 2. If any device retains non-conforming labeling, relabel the device and discard of the incorrect label and application of the corrected labeling provided by Xtant Medical. 3. No product return is required unless a device cannot be confidently verified or corrected on site; please contact Xtant Medical for further guidance in such cases. 4. Complete Correction Acknowledgement to confirm completion of the above actions. For questions, Rebecca Lennemann, SVP, Quality & Regulatory Affairs, at 1-406-924-5878, Monday through Friday, 8 a.m. to 5 p.m. (Mountain Time).
Initiated
2026-02-18
Root cause
Under Investigation by firm
Distribution
US: OR
Recall number
Z-1773-2026
Recall status
Open, Classified
CDRH record ID
219029
FDA classified
2026-04-03
Recall event number
98541

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗