Open original FDA record ↗Recall record
- FDA identifier
- Z-1774-2008
- Company in FDA record
- Medtronic Heart Valve Division
- FDA classification date
- Aug 27, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Medtronic has identified an issue with 685 Hall Easy-Fit valves Models A7700 and M7700 in which the holder of the valve was incorrectly secured to the valve. This could potentially make it difficult to easily disengage the holder from the valve.
Source facts
- Action
- A Medtronic "Urgent Communication" letter dated March 24, 2008 was sent to consignees on 3/25/08. The letter describes the product and problem. The letter requests that they return the impacted devices and to return the reconciliation form "Medtronic Recall Certificate (FCA0807)" confirming whether they have the valves in inventory or if they have been implanted.
- Initiated
- 2008-03-25
- Root cause
- Process control
- Terminated
- 2012-05-15
- Distribution
- Class II Recall - Worldwide Distribution --- including USA states of TX, WY, CA, MI, WI, MO, IL, NY, TN, AR, LA, KY, WA, WV, and countries of Canada, Hong Kong, Australia, Albania, Belgium, Croatia, Germany, India, Italy, Netherlands, Poland, Romania, Russian Federation, South Africa, and Turkey.
- Recall number
- Z-1774-2008
- Recall status
- Terminated
- CDRH record ID
- 70252
- FDA classified
- 2008-08-27
- Recall event number
- 47795