Open original FDA record ↗Recall record
- FDA identifier
- Z-1775-2018
- Company in FDA record
- ICU Medical, Inc.
- FDA record creation date
- May 9, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When continuous cardiac output monitoring with a pulmonary artery catheter is placed in the Pause mode, the catheter thermal coil may continue while in pause mode.
Source facts
- Action
- On about 01/05/2018 customers were notified via an Urgent: Medical Device Correction letter of the recall. The letter identified the affected device and the reason for the correction. Instructions included to inspect current inventory for affected devices and complete and return the Urgent: Medical device Correction Response Form to recall@icumed.com. Customers were notified that they did not need to discontinue using the Cogent Hemodynamic Monitoring System (HMS), but were instructed to disconnect the thermal coil cable from the monitor to discontinue continuous cardiac output (CCO) instead of using the 'Pause' button. Instructions on how to disconnect and reconnect the thermal coil cable from the monitor were provided. Customers may contact ICU Customer Service at 1-866-829-9025, if there are additional questions.
- Initiated
- 2018-01-05
- Root cause
- Software design
- Terminated
- 2019-03-18
- Distribution
- Distributed to the states of CA, FL, NC, NE, NY, and WA.
- Recall number
- Z-1775-2018
- Recall status
- Terminated
- CDRH record ID
- 163535
- FDA record created
- 2018-05-09
- Recall event number
- 79814