Open original FDA record ↗Recall record
- FDA identifier
- Z-1775-2022
- Company in FDA record
- Covidien Llc
- FDA classification date
- Sep 22, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
Source facts
- Action
- Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures. All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to replace the affected ventilator turbine component. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
- Initiated
- 2022-08-19
- Root cause
- Process control
- Distribution
- Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.
- Recall number
- Z-1775-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195688
- FDA classified
- 2022-09-22
- Recall event number
- 90766