Skip to content

Public FDA record

Z-1776-2008: Q-Logic Controller Software used on the following power wheelchairs: Q6000, Q6000XL, Q6000Z, R4000, Q600,

Recall record Z-1776-2008 for Pride Mobility Products Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1776-2008
Company in FDA record
Pride Mobility Products Corp
FDA classification date
Aug 27, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software - Watch dog timer feature was disabled.

Source facts

Action
The recalling firm issued an Urgent Device Correction letter on 4/11/08. This letter instructs the dealers to contact the end users so that the units can be upgraded to enable the watch dog feature.
Initiated
2008-04-11
Root cause
Software design
Terminated
2009-11-13
Distribution
Class II Recall - Worldwide Distribution --- including USA and Canada.
Recall number
Z-1776-2008
Recall status
Terminated
CDRH record ID
70265
FDA classified
2008-08-27
Recall event number
47807

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗