Skip to content

Public FDA record

Z-1776-2019: Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)

Recall record Z-1776-2019 for Conformis, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1776-2019
Company in FDA record
Conformis, Inc.
FDA record creation date
Jun 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.

Source facts

Action
Conformis issued on 29 Mar 2019 via e-mails to surgeons with upcoming surgeries the following: 1. For surgeons who have already performed the surgeries (i.e. for patients whose implant surgeries were completed between July 31st, 2018 and March 27th, 2019), a letter sent letter advising them of the issue. Confirmation of receipt will be requested. 2. For surgeons who have not performed the surgeries, but have approved the iView surgical plan on or before March 27, 2019, and Conformis has been notified of confirmed surgery dates or Conformis has provided commitment dates on or before May 17th 2019, a. They will be provided with a marked-up iView surgical plan that provides the true dimensions of the manufactured implant and asked to choose to (i) continue with surgery, (ii) postpone surgery to allow Conformis to manufacture a new implant with a head/neck length matching the original approved iView or (iii) cancel the order. b. Should the physician choose to proceed with the surgery using the manufactured implant with the 3 mm discrepancy, it will be shipped to them after receiving their written approval provided that the surgery is scheduled on or before May 17, 2019. c. Should the physician choose to postpone surgery, a new implant will be manufactured to match the approved iView and provided to the surgeon
Initiated
2019-03-29
Root cause
Device Design
Terminated
2020-04-28
Distribution
Nationwide
Recall number
Z-1776-2019
Recall status
Terminated
CDRH record ID
172709
FDA record created
2019-06-13
Recall event number
82722

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.