Open original FDA record ↗Recall record
- FDA identifier
- Z-1776-2024
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- May 9, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Source facts
- Action
- On April 10, 2024, Medtronic Navigations, Inc. issued an Urgent: Medical Device Correction to affected consignees via UPS. In addition to informing consignees about the issue, Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. a. Medtronic will work expeditiously with your facility to coordinate a software update at no charge to resolve this anomaly. 3. Pass this notice on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 4. Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic. Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail). Call FDA at (800) FDA-1088. Call Medtronic Technical Services at 1-888-826-5603.
- Initiated
- 2024-03-26
- Root cause
- Software design
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
- Recall number
- Z-1776-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207180
- FDA classified
- 2024-05-09
- Recall event number
- 94411